top of page
4700 TGFX.webp


Stat Fax 4700

ELISA Semi-Automated Microstrip Reader

Programmable open system

 

The Stat Fax 4700 is ideal for reading ELISA microwell strips in laboratories and institutions where small runs of ELISA assays are routine. Assays can easily be programmed using the touch screen interface with filters ranging from 340 nm - 630 nm. The robust instrument is factory calibrated and maintenance free.

  • Economical open system & Compact design

  • Three strip loading capacity

  • Automatically positions each strip for reading

  • Multiple calculation modes

  • Touch screen display and on-board printer

  • Dependable maintenance-free design

  • Up to 120 user programmable stored tests

  • Test results in 30 seconds per 12 well strip

  • Custom filter configurations available

Stat Fax 47001.jpg

Updated interface for easy navigation

Stat Fax 47002.jpg

Simple to use graphically driven utilities

High resolution capacitive touch display

PRODUCT CHARACTERISTICS

  • Linear Measurement Range: 0.0 to 3.0 Absorbance units (A).

  • Photometric Accuracy: +/- (1% +/- 0.010) 0.0 through 1.5 ABS.

                                      +/- (2% +/- 0.010) 1.5 through 3.0 ABS.

  • Stability: Drift of no more than 0.005A in 8 hours/bichromatic.

  • Light Source: Tungsten-Xenon lamp with lamp saver feature.

  • Standard Wavelengths: Standard: 405, 450, 492, and 630 nm

                                        6-VIS: 405 - 630 nm

                                        6-UV: 340 - 630 nm

  • Vessel: Non break apart or break apart strips up to 12 wells long 3 strip loading capacity, (for break apart wells, partial strips or even a single well can be read).

  • Transport mechanism: Auto-Track™ automatic strip transport Detection Mode.

  • Display: Interactive touch screen 3.5” (8 cm) LCD, color graphic display.

  • Printer: Thermal with graphic capability, approximately 29 characters per line.

  • Speed: Reads, calculates, and prints results, in approximately 30 seconds per 12-well strip.

  • Calculation Modes: Single-point calibration by standard or factor, a multi-point calibration with point-to-point curve fit, linear regressions with log, linear, and log-logit selections.

  • Test Storage:120 open channels to store tests Stores all parameters including wavelengths, calculations, unit codes, ranges, interpretations, calibrator values, test names, and previous curve.

  • Dimensions & weight: Approximately 9”x13.5”x5” (24x34x13 cm) weight: 5 lbs (2.3 kg).